portable quality analyzer
Download Data Sheet
Product Code: ATW6200
$0.00

Qty

Ask About This Product
application

ATW6200 is a kind of portable water quality heavy metal rapid analyzer, which adopts electrochemical rapid detection method for analyzing heavy metals. ATW6200 has a fast measurement time, which is less than 5 minutes (the fastest is less than 30s). The sensitivity is high and the detection limit is less than 0.5ppb. The detection accuracy is high and the algorithm design is optimized. The computer software control is 100ppb ≤ ±2% and the stand-alone operation is 100ppb≤±5%.
ATW6200 adopts independent glassy carbon working electrode (corresponding test item: Zn/Cd/Pb/Cu). It has independent gold working electrode (corresponding test item: Hg/As) and independent platinum counter electrode as well as independent silver/silver chloride reference electrode. Measurement standard liquid indication error is less than 3% (@100ppb) and repeatability RSD less than 2%.The design features of ATW6200 include intelligent de-spectrification algorithm, over 24-hour long-term measurement, multiple charging methods, intelligent menu navigation, large-capacity data

  • Fast measurement, measurement time less than 5s.
  • High sensitivity, detection limit less than 0.5ppb
  • Detection for more than 10 kinds of heavy metal
  • Good electrode performance, long life, stable and reliable, easy to maintain
  • Low cost of use, low cost of actual consumables, less consumption, safe and non-toxic
  • High detection accuracy, optimized algorithm design, detection accuracy better than 1ppb
  • Intelligent operation, intelligent guide menu, online operation, support for secondary development
  • Good portability, lightweight suitcase protective packaging, easy to carry
  • Extensible measurement of hexavalent chromium, nickel, manganese and other indicators
  • Rapid detection of heavy metals in water quality by environmental monitoring agencies at all levels, environmental science institutes, universities, water plants, public security inspection departments, enterprises, etc.
  • Applied to on-site emergency detection (such as emergency monitoring of water environment pollution accidents)
  • Laboratory heavy metal detection and analysis
  • Monitoring of surface water, groundwater, seawater, industrial wastewater and drinking water or other environmental water bodies

1.Does the machine cover all the following parameters? Can these detection levels be achieved with clean water?
Here are the metal discharge standards for the treated water (with the ATW6200's measurable range attached based on your last provided range):
Arsenic (As): 1.9 mg/L (Detection Range: 0.001-20 mg/L)
Cadmium (Cd): 0.015 mg/L (Detection Range: 0.0001-20 mg/L)
Hexavalent Chromium (Cr VI): 0.0015 mg/L
Chromium (Cr): 35 mg/L
Cobalt (Co): 0.004 mg/L (Detection Range: 0.0015-15 mg/L)
Copper (Cu): 0.005 mg/L (Detection Range: 0.0001-25 mg/L)
Lead (Pb): 0.005 mg/L (Detection Range: 0.0005-25 mg/L)
Mercury (Hg): 0.000016 mg/L (Detection Range: 0.0005-25 mg/L)
Nickel (Ni): 0.02 mg/L (Detection Range: 0.0015-0.2 mg/L)
Selenium (Se): 0.005 mg/L
Vanadium (V): 0.02 mg/L
Zinc (Zn): 0.05 mg/L (Detection Range: 0.0005-20 mg/L)
Note: No detection range for Selenium, but others can be extended.

2.Does clean water require pre-digestion of samples? Does untreated wastewater require digestion?
No pre-digestion is required.

3.Which electrodes are needed to cover the listed metal parameters?
Separate glassy carbon working electrode (for Zn, Cd, Pb, Cu)
Separate gold working electrode (for Hg, As)
Separate platinum counter electrode
Separate Ag/AgCl reference electrode

4.Can all these metals be read simultaneously, or do some require separate readings?
Requires electrode replacement for measurement.

5.What is the total reading time to cover all listed parameters?
Detection time is less than 5 minutes, with the fastest detection time being less than 30 seconds.

6.Is there any interference from seawater?
There is a certain level of interference.

7.How long can the electrodes be used? What is the replacement cost of the electrodes?
One to two years, with a single electrode costing 800-1200 RMB.

8.Do we need chemicals on our side? Do you have a reagent list that comes with this equipment?
For detecting elements: Cd, Pb, Cu, Zn, Hg, As
Detection Range:
Copper: 0.1 ppb ~ 25 ppm
Lead: 0.5 ppb ~ 25 ppm
Mercury: 0.1 ppb ~ 6 ppm
Cadmium: 0.1 ppb ~ 20 ppm
Zinc: 0.5 ppb ~ 20 ppm
Arsenic: 1 ppb ~ 20 ppm

9.Does high turbidity in water cause interference?

The sample water should not be too turbid, as it will cause interference.