Availability: In Stock
200-1000nm High-sensitivity & High-resolution UV-Vis Spectrometer
Product Code: ATP5030

Qty

Request a Quote or Info
application
Optosky ATP5030 UV-Vis Spectrometer employs the ultra-high performance, high-sensitivity, high resolution, and cost-effective model.  It's ideal for Raman system well. It uses 2048 x 64 pixel, a back-thinned CCD array, TE-cooled down to -10 ℃. Sensor noise is reduced, and SNR is almost 2 times higher than others. Increased reliability can remain measure result do not change with temperature.

Customized design of ultra-low noise CCD signal processing circuit built inside, Noise is lower can 3 counts is ahead of records in the industries. 

ATP5030 is designed with SMA905 optic fiber entrance connector or free space, and it outputs spectral data via USB2.0/UART

Detector
Detector type Back - illum CCD
Spectral range 200-1000 nm
Effective pixels 2048
Pixels dimension 14μm×14μm
Full well capacity 300 ke-
Detector model Hamamtsu s10420+TEC or s11850
Sensitivity 6.5 uV/e-
Dark noise 300 e-rms
Optical parameter
Wavelength range 200-1000nm
Optical resolution 0.03-2 nm
Signal-to-noise >1300:1
Dynamic range 50000:1
Working temperature -10°c to + 50°c
Numberical aperture 0.06
Working humidity < 90% NC
Trigger YES (IN and OUT)
Cooling temperature -5°c
Full Well 300 ke-
Scanning rate < 1 kfps
Optical configuration
Optical design M-type
Focal distance 75mm
Incident slit 50 μm (5, 10, 25, 50, 100, or 200 μm optional)
Incident Interface SMA905 connector,free space
Electrical parameter
Integration time 1 ms - 30 s
Interfaces USB 2.0 (480MHz High speed)
A/D conversion resolution 18 bit
Supply voltage DC4.5 to 5.5 V (type @5V)
Operating current <2.3A
Storage temperature -30°c to + 70 °c
Physical parameters
Dimension 150 × 90 × 37.5 mm^3
Weight 0.5kg
Sealing Anit-sweat

Detector:back-thinned CCD, TE-cooled down to -10 ℃

CCD parameters:2048×64pixel, 14×14um

Ultra-low noise CCD signal processing circuit

Spectral range: 180-1150nm

Spectral resolution: 0.1-2 nm(depend on spectral range, slit width)

Optical path:crossed Czerny-Turner (C-T)

Integration time:2ms-130s

power supply:DC 5V±10% @ <2.3A

18 bit, 570KHz A/D Converter

Entrance connector:SMA905 connector or free space

Output interface: high-speed USB2.0 or UART

20 pins, dual rows programmable extension connector

Raman spectrometer

Micro spectrophotometer, 

high-speed spectrophotometer

Spectral analysis /radiation spectrophotometer/ spectrophotometric analysis

Transmittance, absorbance, reflectance detection

Wavelength detection

LIBS


Q:What is the wavelength range?

 A:ATP5030 is available in 350-800nm,200-850nm,200-1000nm,600-1000nm,800-1000nm

Q:What is the sensitivity?

 A:ATP 5030 6.5 uV/e-

Q:What is the resolution?

 A:Resolution of ATP5030 can be customized between 0.03-2 nm corresponding to the different slits and spectral range.

Q:What is signal-to-noise ratio?

 A:SNR of ATP5030 is   >1300:1.

Q:What brands used for the detector?

 A:CCD of ATP5030used Hamamatsu made in Japan, the best brand of CCD.

Q:What is the size?

 A:150mm×90mm×37.5mm

Q:What is the weight?

 A:0.55kg.

Q:Do you have the software interface for C++, Labviewer?

 A:ATP 5030 is equipped with SDK software for C++ , labviewr development.

Q:Do you have the software interface for Matlab development?

 A:Not at the moment, if the customer has a bigger volume to support development, that's fine.

Q:How long is the spectrometer warranty?

 A:Leading industrial warranty is three year, and OEM project is offered a one-year warranty.

Q:What is the price?

 A:Contact us and inform spectral range, slit size, grating and resolution for a quote.

Write a review